VOLTAGE |
RANGES Frequency Range Peak Continuous Peak < 1 second Input Impedance Display Crest Factor |
900, 215, 46, 10 Vpk 10Hz to 1MHz 1500 Vpk (over voltage) 5000 Vpk (over voltage) 1 M¥Ø 4.5 Digits 20 (Peak / RMS) |
RMS 45-65Hz |
Accuracy |
0.05% of rdg +0.05% of range |
RMS |
Accuracy |
0/1% of rdg + 0.1% of range + 4mV + (0.02 * F)% of rdg |
DC |
Accuracy |
0.1% of rdg + 0.4% of range + 5mV |
VOLTAGE +/- PEAK |
Accuracy |
0.5% of rdg + 0.5% of range + (0.02 * F)% of rdg |
CURRENT |
RANGES Frequency range Peak Continuous Peak < 1 second Input resistance Crest Factor |
100, 25, 6.25, 1.6, 0.4, 0.1 Apk 10Hz to 1MHz 20Arms 60Arms (over current) 12.5m¥Ø 20 (Peak/RMS) |
RMS 45-65Hz |
Accuracy |
0.1% of rdg +0.1% of range |
RMS |
Accuracy |
0.1% of rdg + 0.1% of range + 1mA + (0.02 * F)% of rdg |
DC |
Accuracy |
0.1% of rdg + 0.4% of range + 1mA |
CURRENT +/- PEAK |
Accuracy |
0.5% of rdg + 0.5% of range + (0.02 * F)% of rdg |
WATTS |
RANGES Frequency range |
1W to 90kW 10Hz to 1MHz |
RMS 45-65Hz |
Accuracy |
[{(Verror/Vreading) +(Aerror/Areading)}*Wreading + 0.1% of W reading |
|
Accuracy |
0.2% rdg + 0.1% range + 4mW + ((0.05/PF) * F)% of rdg |
VA |
RANGES Frequency range Accuracy |
1VA to 90kVA 10Hz to 1MHz 0.2% rdg + 0.1 % of range +4mVA + (0.05 * F)% of rdg |
VAr |
RANGES Frequency range Accuracy |
1VAr to 90kVAr 10Hz to 1MHz 0.2% rdg + 0.1% range + 4mVAr +((0.05/1-PF) * F)% of rdg |
POWER FACTOR |
Power Factor Range Accuracy |
-1.000 to +1.000 +/-0.002 +/- ((0.001/PF) * F) |
FREQUENCY |
RANGE Accuracy |
DC and 10Hz to 1MHz 0.1% |
VOLTAGE CREST FACTOR |
RANGE Accuracy |
1.00 to 20.0 %Vpk error + % Vrms error |
CURRENT CREST FACTOR |
RANGE Accuracy |
1.00 to 20.0 %Apk error + % Arms error |
PEAK INRUSH CURRENT |
RANGE Accuracy |
100Apk 2% of range +/- 20mA |
HARMONIC ANAYLSIS |
Number of Voltage & Current Harmonics Maximum Harmonics Frequency Accuracy
Frequency Range |
50
450kHz 0.2% of Reading + 0.1% of range +0.04% per kHz of Harmonics 10Hz to 450kHz |
THD Total Harmonic Distortion |
Range & Accuracy
Formula |
Range 0-999% Accuracy 0.4% + (0.1 * F)% of reading Series or difference |
STANDBY POWER |
Time Window Resolution |
1-300 sec 1 second |
IMPEDANCE |
Range Accuracy |
0.005¥Ø to 1M¥Ø 0.2% of Reading +0.1% of range +5m¥Ø + ((0.05/PF) * F)% of reading |
RESISTANCE |
Range Accuracy |
0.005¥Ø to 1M¥Ø 0.2% of Reading + 0.1% of range +5m¥Ø + ((0.05/1-PF) * F)% of reading |
REACTANCE |
Range Accuracy |
0.005¥Ø to 1M¥Ø 0.2% of reading + 0.1% of range +5m¥Ø + ((0.05/PF) * F)% of reading |
EXTERNAL SHUNT Scaling |
Input Range |
+/- 1250 mVpk 0.0001 to 100000 |
MECHANICAL |
|
1/2 rack size. Rack height 85mm W = 224mm. Height including feet 103mm D = 285mm. Weight 3.21Kg (7lbs.) |